Abstract
Although ultrasound elastography is established as a reliable and valid tool for assessment of skeletal muscles, guidelines around the technical specifications, data selection, and acquisition parameters still lack consensus. One such parameter is the use of the quantification box (Q-box) that calculates the shear wave velocity/modulus, within a selected region of interest (ROI). Currently, no data compare the effect of the elastographic area within the ROI to the mean shear wave velocity calculations, using a Q-box. In this study, the mean shear wave velocity calculated over a smaller (single Q-box) ROI is compared to the mean shear wave velocity calculated over maximum area of elastogram, within a ROI. Comparison of mean shear wave velocity revealed a significant difference (t = 2.79, P =.007) between the means calculated over maximum area of elastogram for only nonuniform elastograms. The rater agreement for the classification scheme was assessed (κ = 0.85). To prevent possible overestimation of shear wave velocities, it may be necessary to place the Q-box over the maximum elastographic area.