Skip to main navigation Skip to search Skip to main content

Die korrektur des hypoplastischen aortenbogens unter anwendung der kombinierten myokard-und zerebralperfusion

Translated title of the contribution: Correction of the hypoplastic aortic arch using combined myocardial and cerebral perfusion
  • Anika Villanueva
  • , J. Meuser
  • , J. Gehron
  • , S. Nisztuk
  • , C. Yerebakan
  • , K. Valeske
  • , H. Elmontaser
  • , M. Müller
  • , H. Akintürk

Research output: Contribution to journal › Article › peer-review

Abstract

During surgical correction of the hypoplastic aortic arch, the protection of end organs is of pivotal importance as clamping of the aorta does not allow simultaneous perfusion of all end organs. Correction under deep hypothermic circulatory arrest is limited by the time of circulatory arrest and increases the corresponding morbidity. Selective cerebral perfusion with flow rates around 20–30 % of normal flow at 28 °C makes circulatory arrest redundant. The still present myocardial ischemia may be prevented through additional and simultaneous myocardial perfusion. Correction of the hypoplastic aortic arch is conducted with simultaneous cerebral and myocardial perfusion at 28–30 °C. Both compartments are simultaneously perfused through the arterial line. With an innominate artery cannula of 6–8 F and a 4 F aortic root cannula, the overall flow of simultaneous cerebral and myocardial perfusion accounts for 34.8–48.4 % (38.9 ± 5 %) of full flow. Myocardial perfusion showed stable flow of around 8.6–10.5 % (9.8 ± 0.6 %) of full flow. In contrast to this, the ratio of selective myocardial and cerebral perfusion showed a higher variation between 27.9– 39.5 % (34.3 ± 4.1 %). Complex isolated perfusion of different vascular compartments is possible in clinical practice. The particular corresponding flow rates of the y-configuration between the cerebral and myocardial compartments are primarily determined by the geometric ratio of the cross section of both cannulae. Nevertheless, resistance changes within a single compartment may result in a marked flow increase in the remaining vascular bed. Unfortunately flow rates of isolated vascular compartments have not been defined so far. As disturbing flow changes cannot be prevented, complex isolated perfusion should only be carried out with near-infrared spec troscopy and careful monitoring and venting of the beating myocardium.

Translated title of the contributionCorrection of the hypoplastic aortic arch using combined myocardial and cerebral perfusion
Original languageGerman
Pages (from-to)66-69
Number of pages4
JournalKardiotechnik
Volume24
Issue number3
StatePublished - Sep 2015

Keywords

  • Cerebral circulation
  • Coarctatio aortae
  • Myocardial circulation
  • Neonates
  • Perfusion methods