The current clinical interface for Varian's intrafraction motion review (IMR) is limited, providing only qualitative data for review at the treatment console. This study provides a method of extracting and interpreting data from combined log files for quantitative evaluation. Combined log files acquired during patient treatment and a parsing code was developed to scan the combined log file looking for unique identifiers pertaining to the data of interest. We were able to extract clinically relevant parameters from the log files including date and time, gantry angle, expected marker position, found marker position, pixel size, and detection result. This study details how to compare IMR data to Calypso investigating dual-surrogates for intrafraction monitoring during treatment for other researchers to build on these methods. Understanding data recorded during treatment within the combined log files can be helpful in quality improvement of patient care by retrospectively reviewing intrafraction motion.

Original languageEnglish
Pages (from-to)196-202
Number of pages7
JournalJournal of Applied Clinical Medical Physics
Volume22
Issue number11
DOIs
StatePublished - Nov 2021

Keywords

  • Motion management (intrafraction)
  • dual surrogate
  • x-ray projection

Fingerprint

Dive into the research topics of 'Interpretation of intrafraction motion review data and method for verification'. Together they form a unique fingerprint.