T1 - Intraoperative detection of 18F-FDG-avid tissue sites using the increased probe counting efficiency of the K-alpha probe design and variance-based statistical analysis with the three-sigma criteria.
AU - Povoski, Stephen P.
AU - Chapman, Gregg J.
AU - Murrey, Douglas A.
AU - Lee, Robert
AU - Martin, Edward W.
AU - Hall, Nathan C.
PY - 2013
Y1 - 2013
N2 - Intraoperative detection of (18)F-FDG-avid tissue sites during 18F-FDG-directed surgery can be very challenging when utilizing gamma detection probes that rely on a fixed target-to-background (T/B) ratio (ratiometric threshold) for determination of probe positivity. The purpose of our study was to evaluate the counting efficiency and the success rate of in situ intraoperative detection of (18)F-FDG-avid tissue sites (using the three-sigma statistical threshold criteria method and the ratiometric threshold criteria method) for three different gamma detection probe systems. Of 58 patients undergoing (18)F-FDG-directed surgery for known or suspected malignancy using gamma detection probes, we identified nine (18)F-FDG-avid tissue sites (from amongst seven patients) that were seen on same-day preoperative diagnostic PET/CT imaging, and for which each (18)F-FDG-avid tissue site underwent attempted in situ intraoperative detection concurrently using three gamma detection probe systems (K-alpha probe, and two commercially-available PET-probe systems), and then were subsequently surgical excised. The mean relative probe counting efficiency ratio was 6.9 (± 4.4, range 2.2-15.4) for the K-alpha probe, as compared to 1.5 (± 0.3, range 1.0-2.1) and 1.0 (± 0, range 1.0-1.0), respectively, for two commercially-available PET-probe systems (P < 0.001). Successful in situ intraoperative detection of 18F-FDG-avid tissue sites was more frequently accomplished with each of the three gamma detection probes tested by using the three-sigma statistical threshold criteria method than by using the ratiometric threshold criteria method, specifically with the three-sigma statistical threshold criteria method being significantly better than the ratiometric threshold criteria method for determining probe positivity for the K-alpha probe (P = 0.05). Our results suggest that the improved probe counting efficiency of the K-alpha probe design used in conjunction with the three-sigma statistical threshold criteria method can allow for improved detection of 18F-FDG-avid tissue sites when a low in situ T/B ratio is encountered.
AB - Intraoperative detection of (18)F-FDG-avid tissue sites during 18F-FDG-directed surgery can be very challenging when utilizing gamma detection probes that rely on a fixed target-to-background (T/B) ratio (ratiometric threshold) for determination of probe positivity. The purpose of our study was to evaluate the counting efficiency and the success rate of in situ intraoperative detection of (18)F-FDG-avid tissue sites (using the three-sigma statistical threshold criteria method and the ratiometric threshold criteria method) for three different gamma detection probe systems. Of 58 patients undergoing (18)F-FDG-directed surgery for known or suspected malignancy using gamma detection probes, we identified nine (18)F-FDG-avid tissue sites (from amongst seven patients) that were seen on same-day preoperative diagnostic PET/CT imaging, and for which each (18)F-FDG-avid tissue site underwent attempted in situ intraoperative detection concurrently using three gamma detection probe systems (K-alpha probe, and two commercially-available PET-probe systems), and then were subsequently surgical excised. The mean relative probe counting efficiency ratio was 6.9 (± 4.4, range 2.2-15.4) for the K-alpha probe, as compared to 1.5 (± 0.3, range 1.0-2.1) and 1.0 (± 0, range 1.0-1.0), respectively, for two commercially-available PET-probe systems (P < 0.001). Successful in situ intraoperative detection of 18F-FDG-avid tissue sites was more frequently accomplished with each of the three gamma detection probes tested by using the three-sigma statistical threshold criteria method than by using the ratiometric threshold criteria method, specifically with the three-sigma statistical threshold criteria method being significantly better than the ratiometric threshold criteria method for determining probe positivity for the K-alpha probe (P = 0.05). Our results suggest that the improved probe counting efficiency of the K-alpha probe design used in conjunction with the three-sigma statistical threshold criteria method can allow for improved detection of 18F-FDG-avid tissue sites when a low in situ T/B ratio is encountered.
UR - http://www.scopus.com/inward/record.url?scp=84874443173&partnerID=8YFLogxK
U2 - 10.1186/1471-2407-13-98
DO - 10.1186/1471-2407-13-98
M3 - Article
C2 - 23496877
AN - SCOPUS:84874443173
VL - 13
JO - Unknown Journal
JF - Unknown Journal
M1 - 98