Skip to main navigation Skip to search Skip to main content

Methotrexate and 7-hydroxymethotrexate: Serum level monitoring by high-performance liquid chromatography

  • C. Canfell
  • , Wolfgang Sadee

Research output: Contribution to journalArticlepeer-review

Abstract

We have developed a reverse-phase high-performance liquid chromatographic assay with ultraviolet detection at 303 nm for methotrexate (MTX) and 7-hydroxymethotrexate (7-OH-MTX) in serum, plasma, and cerebrospinal fluid. The procedure involves the concentration of MTX and 7-OH-MTX in the aqueous phase by an n-butanol-ether extraction, resulting in a sensitivity of 10-20 nM in serum. Serum levels of MTX correlated well with those obtained by a competitive dihydrofolate reductase binding assay (r = 0.989). The clinical utility of the assay was demonstrated by measuring MTX and 7-OH-MTX serum levels in 42 patient samples.

Original languageEnglish
Pages (from-to)165-169
Number of pages5
JournalCancer Treatment Reports
Volume64
Issue number1
StatePublished - Jan 1 1980

Fingerprint

Dive into the research topics of 'Methotrexate and 7-hydroxymethotrexate: Serum level monitoring by high-performance liquid chromatography'. Together they form a unique fingerprint.