TY - JOUR
T1 - Monitoring Pulmonary Arterial Hypertension Using an Implantable Hemodynamic Sensor
AU - Benza, Raymond L.
AU - Doyle, Mark
AU - Lasorda, David
AU - Parikh, Kishan S.
AU - Correa-Jaque, Priscilla
AU - Badie, Nima
AU - Ginn, Greg
AU - Airhart, Sophia
AU - Franco, Veronica
AU - Kanwar, Manreet K.
AU - Murali, Srinivas
AU - Raina, Amresh
AU - Agarwal, Rahul
AU - Rajagopal, Sudarshan
AU - White, Jason
AU - Biederman, Robert
N1 - Funding Information:
FUNDING/SUPPORT: The work performed at Allegheny General Hospital was supported by a National Heart, Lung, and Blood Institute VITA Contract [Contract HHSN268201400008C], and the work performed at Duke Clinical Research Institute was funded by an investigator-initiated award from Abbott.
Publisher Copyright:
© 2019 American College of Chest Physicians
PY - 2019/12
Y1 - 2019/12
N2 - Background: Pulmonary arterial hypertension (PAH) is a chronic disease that ultimately progresses to right-sided heart failure (HF) and death. Close monitoring of pulmonary artery pressure (PAP) and right ventricular (RV) function allows clinicians to appropriately guide therapy. However, the burden of commonly used methods to assess RV hemodynamics, such as right heart catheterization, precludes frequent monitoring. The CardioMEMS HF System (Abbott) is an ambulatory implantable hemodynamic monitor, previously only used in patients with New York Heart Association (NYHA) class III HF. In this study, we evaluate the feasibility and early safety of monitoring patients with PAH and right-sided HF using the CardioMEMS HF System. Methods: The CardioMEMS HF sensors were implanted in 26 patients with PAH with NYHA class III or IV right-sided HF (51.3 ± 18.3 years of age, 92% women, 81% NYHA class III). PAH therapy was tracked using a minimum of weekly reviews of CardioMEMS HF daily hemodynamic measurements. Safety, functional response, and hemodynamic response were tracked up to 4 years with in-clinic follow-ups. Results: The CardioMEMS HF System was safely used to monitor PAH therapy, with no device-related serious adverse events observed and a single preimplant serious adverse event. Significant PAP reduction and cardiac output elevation were observed as early as 1 month postimplant using trends of CardioMEMS HF data, coupled with significant NYHA class and quality of life improvements within 1 year. Conclusions: The CardioMEMS HF System provided useful information to monitor PAH therapy, and demonstrated short- and long-term safety. Larger clinical trials are needed before its widespread use to guide therapy in patients with severe PAH with right-sided HF.
AB - Background: Pulmonary arterial hypertension (PAH) is a chronic disease that ultimately progresses to right-sided heart failure (HF) and death. Close monitoring of pulmonary artery pressure (PAP) and right ventricular (RV) function allows clinicians to appropriately guide therapy. However, the burden of commonly used methods to assess RV hemodynamics, such as right heart catheterization, precludes frequent monitoring. The CardioMEMS HF System (Abbott) is an ambulatory implantable hemodynamic monitor, previously only used in patients with New York Heart Association (NYHA) class III HF. In this study, we evaluate the feasibility and early safety of monitoring patients with PAH and right-sided HF using the CardioMEMS HF System. Methods: The CardioMEMS HF sensors were implanted in 26 patients with PAH with NYHA class III or IV right-sided HF (51.3 ± 18.3 years of age, 92% women, 81% NYHA class III). PAH therapy was tracked using a minimum of weekly reviews of CardioMEMS HF daily hemodynamic measurements. Safety, functional response, and hemodynamic response were tracked up to 4 years with in-clinic follow-ups. Results: The CardioMEMS HF System was safely used to monitor PAH therapy, with no device-related serious adverse events observed and a single preimplant serious adverse event. Significant PAP reduction and cardiac output elevation were observed as early as 1 month postimplant using trends of CardioMEMS HF data, coupled with significant NYHA class and quality of life improvements within 1 year. Conclusions: The CardioMEMS HF System provided useful information to monitor PAH therapy, and demonstrated short- and long-term safety. Larger clinical trials are needed before its widespread use to guide therapy in patients with severe PAH with right-sided HF.
KW - CardioMEMS
KW - hemodynamics
KW - pulmonary hypertension
KW - ride-sided heart failure
UR - https://www.scopus.com/pages/publications/85072162792
U2 - 10.1016/j.chest.2019.06.010
DO - 10.1016/j.chest.2019.06.010
M3 - Article
C2 - 31265832
AN - SCOPUS:85072162792
SN - 0012-3692
VL - 156
SP - 1176
EP - 1186
JO - Chest
JF - Chest
IS - 6
ER -