Abstract
The use of fluoroscopic visualization during spinal procedures may subject both the patient and the surgical team to considerable amounts of X-rays. A number of factors are known to influence the radiation doses administered to these individuals, many of which are related to the specific settings of the C-arm such as beam energy, collimation, and tube current, whereas others involve external variables including proper positioning of the image intensifier, protective shielding, and the implementation of appropriate monitoring protocols. By incorporating these safety measures, it may be possible to achieve significant reductions in the exposures of the patient and surgeon alike.
| Original language | English |
|---|---|
| Pages (from-to) | 181-185 |
| Number of pages | 5 |
| Journal | Seminars in Spine Surgery |
| Volume | 20 |
| Issue number | 3 |
| DOIs | |
| State | Published - Sep 2008 |
Keywords
- exposure
- fluoroscopy
- intraoperative
- minimize
- radiation
- spine
Fingerprint
Dive into the research topics of 'Techniques to Minimize Intraoperative Radiation Exposure'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver