Skip to main navigation Skip to search Skip to main content

Theoretical model for an MRI radio frequency resonator

  • Brian A. Baertlein
  • , Özlem Özbay
  • , Tamer Ibrahim
  • , Robert Lee
  • , Ying Yu
  • , Allahyar Kangarlu
  • , Pierre Marie L. Robitaille

Research output: Contribution to journalArticlepeer-review

Abstract

A theoretical model is described for a magnetic resonance imaging (MRI) radio-frequency resonator (an MRI 'coil') that is useful at ultrahigh frequencies. The device is a 'TEM resonator,' which is based on a concept originally proposed by Roschmann (1988). The coil comprises a circular cavity-like structure containing several coaxial transmission-lines operating in a transverse electromagnetic (TEM) mode. The model developed herein treats the empty coil and is based on multiconductor transmission line theory. This work generalizes and extends similiar analyses of the device by Roschmann (1995) and Chingas and Zhang (1996). The model employs explicit calculation of per-unit-length parameters for TEM lines having arbitrary geometries. Calculations of the resonator's frequency response are found to compare well with measurements. Fields produced by linear (single-point) and quadrature drive are also computed and compared to images of low-permittivity phantoms.

Original languageEnglish
Pages (from-to)535-546
Number of pages12
JournalIEEE Transactions on Biomedical Engineering
Volume47
Issue number4
DOIs
StatePublished - 2000

Keywords

  • Magnetic resonance imaging (MRI)
  • Radio frequency (RF) resonator

Fingerprint

Dive into the research topics of 'Theoretical model for an MRI radio frequency resonator'. Together they form a unique fingerprint.